ELISA Reader

TECAN SUNRISE

Optical system

12 measurement channels, 1 reference channel

Wavelength range

340 to 750 nm

Wavelength selection

Interference filters (slide of 4 to 6 filters), Bandwidth: 10 nm

Measurement range

340 to 399 nm: 0 to 3.0 OD, 400 to 750 nm: 0 to 4.0 OD

Measurement time

Single wavelength: 6 seconds
Dual wavelength: 8 seconds
Kinetic repeat time: 5 seconds

Accuracy

 

0.0 to 2.0 OD/492 nm:

better than ± (1.0% + 0.010 OD)

2.0 to 3.0 OD/492 nm:

better than ± (1.5% + 0.010 OD)

 

Linearity

 

340 to 399 nm:

0.0 to 2.0 OD better than ±2%

400 to 750 nm:

0.0 to 2.0 OD better than ±1%
2.0 to 3.0 OD better than ±1.5%

 

Precision

 

0.0 to 2.0 OD/492 nm:

better than ± (0.5% + 0.005 OD)

2.0 to 3.0 OD/492 nm:

better than ± (1.0% + 0.005 OD)

 

Resolution

0.001 OD

Shaking

Linear shaking in 4 different modes